RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Medical Division of Stryker Corporation recalls Medi-Therm Hyper/Hypothermia System, MTA7900

Recalled May 17, 2017 · FDA recall Z-2058-2017 · Stryker Medical Division of Stryker

Hazard
The water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode. · FDA Class II
Units
119
Sold at
USA (nationwide) Distribution and Internationally to Colombia.

Description

The water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode.

Product
Medi-Therm Hyper/Hypothermia System, MTA7900
Manufacturer(s)
Stryker Medical Division of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Medical Division of Stryker Corporation recalls Medi-Therm Hyper/Hypothermia System, MTA7900 | RecallWatchUSA