Stryker Medical Division of Stryker Corporation recalls Model FL14E Rose BedsThe Rose Bed is an AC-powered adjustable long-term care bed intended for medical purposes that co

Recalled November 27, 2013 · FDA recall Z-0363-2014 · Stryker Medical Division of Stryker

Hazard
It was reported that the siderail may appear latched when it is not, due to lack of proper preventive maintenance. This might result in bruises and/or lacerations to a patient who is leaning against the siderails if the siderail releases. Additionally, the siderail may not effectively constrain a patient to prevent them from rolling or falling off the bed. · FDA Class II
Units
4126 units USA; 20449 inits Internationally
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to Canada; Japan, South America, Latin America and China.

Description

It was reported that the siderail may appear latched when it is not, due to lack of proper preventive maintenance. This might result in bruises and/or lacerations to a patient who is leaning against the siderails if the siderail releases. Additionally, the siderail may not effectively constrain a patient to prevent them from rolling or falling off the bed.

Product
Model FL14E Rose BedsThe Rose Bed is an AC-powered adjustable long-term care bed intended for medical purposes that consists of a bed with a built-in motor and controls that can be operated to adjust the height and surface contour of the bed. The device includes movable and latchable siderails
Manufacturer(s)
Stryker Medical Division of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Medical Division of Stryker Corporation recalls Model FL14E Rose BedsThe Rose Bed is an AC-powered adjustable long-term care bed intended for medical purposes that co | RecallWatchUSA