Stryker Medical Division of Stryker Corporation recalls Power-PRO 2- A powered ambulance cot that consists of a platform, including a mattress, mounted on a wheeled, retractabl

Recalled July 19, 2023 · FDA recall Z-2130-2023 · Stryker Medical Division of Stryker

Hazard
Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion. · FDA Class II
Units
1,980 cots
Sold at
Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Switzerland, United Kingdom.

Description

Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.

Product
Power-PRO 2- A powered ambulance cot that consists of a platform, including a mattress, mounted on a wheeled, retractable X-frame that is designed to support and transport a maximum weight of 700 lb (318 kg). Model Number: 6507
Manufacturer(s)
Stryker Medical Division of Stryker Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov