Stryker Medical Division of Stryker Corporation recalls Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Mo
Recalled March 22, 2023 · FDA recall Z-1256-2023 · Stryker Medical Division of Stryker
Hazard
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport. · FDA Class II
Units
4 units
Sold at
US Nationwide distribution in the states of NY, VA.
Description
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.
Product
Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001
Manufacturer(s)
Stryker Medical Division of Stryker Corporation