Home / Recalls / Medical Devices
Stryker Medical Division of Stryker Corporation recalls Stryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered ambulance cots are desig
Recalled June 22, 2016 · FDA recall Z-1990-2016 · Stryker Medical Division of Stryker
Hazard
Complaints of smoke coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In addition to the alleged smoking, one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later i · FDA Class II
Units
770
Sold at
Distributed US (nationwide) and the countries of Australia, Canada, Hong Kong, Spain, Switzerland, and United Kingdom.
Description
Complaints of smoke coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In addition to the alleged smoking, one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later i
Product
Stryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered ambulance cots are designed to support and transport a maximum weight of 700 lb. (318 kg) in pre-hospital and hospital environments.
Manufacturer(s)
Stryker Medical Division of Stryker Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →