Stryker Neurovascular recalls Trevo ProVue, Catalog: 90184

Recalled June 5, 2024 · FDA recall Z-1903-2024 · Stryker Neurovascular

Hazard
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis. · FDA Class II
Units
280
Sold at
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

Description

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

Product
Trevo ProVue, Catalog: 90184
Manufacturer(s)
Stryker Neurovascular

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov