Stryker Orthobiologics, Inc. recalls Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single use device des

Recalled November 28, 2012 · FDA recall Z-0431-2013 · Stryker Orthobiologics

Hazard
When preparing plasma using VitaPrep, it was reported that a small amount of blood collected in the bottom of the centrifuge cup following the centrifugation step. Firm has since discovered that if blood is present in the Luer connection of the VitaPrep following the blood draw step, this blood could be displaced to the outside of the syringe when the VitaPrep cap is reattached to the syringe. T · FDA Class II
Units
3,268
Sold at
Nationwide Distribution-including the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV

Description

When preparing plasma using VitaPrep, it was reported that a small amount of blood collected in the bottom of the centrifuge cup following the centrifugation step. Firm has since discovered that if blood is present in the Luer connection of the VitaPrep following the blood draw step, this blood could be displaced to the outside of the syringe when the VitaPrep cap is reattached to the syringe. T

Product
Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single use device designed to prepare plasma from the patient's blood using a tabletop centrifuge.
Manufacturer(s)
Stryker Orthobiologics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov