Recalled August 16, 2023 · FDA recall Z-2394-2023 · Stryker Sustainability Solutions

Hazard
39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury. · FDA Class II
Units
340
Sold at
US: WI, ID, CA, FL, HI, MD, RI, AZ, MA, GA, IL

Description

39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury.

Product
HoverMatt, 39" W x 78" L, REF: HM39SPU-B
Manufacturer(s)
Stryker Sustainability Solutions

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov