RecallWatchUSA

Home / Recalls / Medical Devices

Stryker Sustainability Solutions recalls Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be used to abrade,

Recalled March 13, 2019 · FDA recall Z-0973-2019 · Stryker Sustainability Solutions

Hazard
Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used. · FDA Class II
Units
14,265 units
Sold at
Nationwide Distribution: AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. OUS Israel and Canada.

Description

Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used.

Product
Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be used to abrade, cut and excise tissue and bone; remove loose fragments; and, shave away debris in arthroscopic surgeries, as well as surgeries of the jaw and sinuses. Intended for use in orthopedic surgical procedures of the joints, jaw or sinuses where the cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.
Manufacturer(s)
Stryker Sustainability Solutions

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Stryker Sustainability Solutions recalls Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be used to abrade, | RecallWatchUSA