Home / Recalls / Medical Devices
SunMed Holdings, LLC recalls Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resus
Recalled November 29, 2023 · FDA recall Z-0369-2024 · SunMed Holdings
Hazard
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing. · FDA Class II
Units
N/A
Sold at
US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Description
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Product
Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant Resuscitator REF #s: CPRM3312FM, CPRM3312FM123, CPRM3312FPM, CPRM3312FPM3, CPRM3312FPMC, CPRM3312M, CPRM3312M3, CPRM3312P3M3, CPRM3312PM, CPRM3312PM2, CPRM3312PMC2, CPRM3312PMTD1, CPRM3312PMX, CPRM3312PWM, CPRM3322FM, CPRM3322M, CPRM3322PM; 3. Pediatric Resuscitator REF #s: CPRM2216FM, CPRM2216FM3, CPRM2216FPM, CPRM2216FPM3, CPRM2216FPMC, CPRM2216M, CPRM2216P3M3, CPRM2216PCM, CPRM2216PM, CPRM2216PM6, CPRM2226FM, CPRM2226FM3, CPRM2226FPM, CPRM2226FPWM, CPRM2226M; 4. Small Adult Resuscitator REF #s: CPRM4416FPMC, CPRM4416M, CPRM4426M.
Manufacturer(s)
SunMed Holdings, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →