RecallWatchUSA

Home / Recalls / Medical Devices

Sunquest Information Systems, Inc. recalls Sunquest Encompass The reporting site is using the Encompass software in an acute patient care setting.

Recalled August 29, 2012 · FDA recall Z-2243-2012 · Sunquest Information Systems

Hazard
Sunquest is recalling the Sunquest Encompass versions 2.4 and later because the specimen collect time is displaying and printing one hour later than the actual collect time only when Daylight Saving Time is active. · FDA Class III
Units
23 clients
Sold at
Worldwide distribution: USA (nationwide) and country of: Bahamas.

Description

Sunquest is recalling the Sunquest Encompass versions 2.4 and later because the specimen collect time is displaying and printing one hour later than the actual collect time only when Daylight Saving Time is active.

Product
Sunquest Encompass The reporting site is using the Encompass software in an acute patient care setting.
Manufacturer(s)
Sunquest Information Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Sunquest Information Systems, Inc. recalls Sunquest Encompass The reporting site is using the Encompass software in an acute patient care setting. | RecallWatchUSA