Home / Recalls / Medical Devices
Superstat Corp recalls Superstat Modified Collagen Hemostatic Sponge, 32mm Disc, 1.5%, Part# 9120-15, 9100-15. For use when hemostasis is de
Recalled November 28, 2012 · FDA recall Z-0339-2013 · Superstat
Hazard
The recall was initiated because Superstat Corporation has confirmed that no documentation are available to verify that quarterly dose sterility audits were performed for the Superstat Modified Collagen Hemostatic Sponge. · FDA Class II
Units
26,422 units
Sold at
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, Thailand, Mexico, Canada, Pakistan, Israel, Greece, Serbia, Canda, Isle of Man, Nova Scotia, and Singapore.
Description
The recall was initiated because Superstat Corporation has confirmed that no documentation are available to verify that quarterly dose sterility audits were performed for the Superstat Modified Collagen Hemostatic Sponge.
Product
Superstat Modified Collagen Hemostatic Sponge, 32mm Disc, 1.5%, Part# 9120-15, 9100-15. For use when hemostasis is desired along suture lines, in diffusely bleeding sites, in incisions, dissections, and around arterial or venous anastomoses. Indicated in oozing bleeding, either capillary or venous, in almost all types of incisions or wounds, and as an adjunct to conventional closure techniques
Manufacturer(s)
Superstat Corp
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →