RecallWatchUSA

Home / Recalls / Medical Devices

Symbios Medical Products, LLC recalls Part510447 Disposable Infusion Pump, ...Elastomeric Pump 600 mL, 2mL/hr/side,... found in the following kits GoPump

Recalled June 12, 2013 · FDA recall Z-1456-2013 · Symbios Medical Products

Hazard
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the · FDA Class I
Units
34517 for all kits
Sold at
US Nationwide distribution.

Description

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Product
Part510447 Disposable Infusion Pump, ...Elastomeric Pump 600 mL, 2mL/hr/side,... found in the following kits GoPump Kit 510462-BP, 5 " catheter & BIOPATCH GoPump Kit 510558-BP 10 " catheter& BIOPATCH GoPump Kit 64310 Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.
Manufacturer(s)
Symbios Medical Products, LLC