Home / Recalls / Medical Devices
Synaptive Medical Inc recalls BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can
Recalled December 5, 2018 · FDA recall Z-0470-2019 · Synaptive Medical
Hazard
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure. · FDA Class I
Units
37 units
Sold at
Worldwide - US Nationwide Distribution and the country of Canada
Description
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.
Product
BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.
Manufacturer(s)
Synaptive Medical Inc
Made in
Canada
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →