Synthes USA HQ, Inc. recalls Synthes Matrix Threaded Persuader Deformity is a comprehensive thoracolumbar pedical screw system designed to provide

Recalled December 4, 2013 · FDA recall Z-0384-2014 · Synthes USA HQ

Hazard
Synthes became aware that during a procedure using the Matrix Threaded Persuader, the Reduction Insert malfunctioned causing pieces to break off. Once there is a breakage to the Reduction Insert it will not allow the instrument to function as required. · FDA Class II
Units
634
Sold at
Worldwide distribution: US (nationwide) and countries of Canada and Switzerland.

Description

Synthes became aware that during a procedure using the Matrix Threaded Persuader, the Reduction Insert malfunctioned causing pieces to break off. Once there is a breakage to the Reduction Insert it will not allow the instrument to function as required.

Product
Synthes Matrix Threaded Persuader Deformity is a comprehensive thoracolumbar pedical screw system designed to provide flexibility biomechanical performance and a solution to complex posterior pathological challenges.
Manufacturer(s)
Synthes USA HQ, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov