RecallWatchUSA

Home / Recalls / Medical Devices

Synthes USA HQ, Inc. recalls Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is

Recalled December 11, 2013 · FDA recall Z-0455-2014 · Synthes USA HQ

Hazard
A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw. · FDA Class II
Units
1
Sold at
Nationwide Distribution - only AZ .

Description

A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw.

Product
Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1.
Manufacturer(s)
Synthes USA HQ, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Synthes USA HQ, Inc. recalls Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is | RecallWatchUSA