Home / Recalls / Medical Devices
Synthes (USA) Products LLC recalls Depuy Synthes Trauma- Trochanteric Fixation Nail Advanced (TFNA) Blade/Screw Guide Sleeve, Surgical Instrument Part Num
Recalled July 27, 2016 · FDA recall Z-2214-2016 · Synthes (USA) Products
Hazard
DePuy Synthes is initiating a voluntary medical device recall due to Blade/Screw Guide Sleeves (Part # 03.037.017) may have been manufactured with a slightly larger than specified outer diameter which could prevent the sleeve from going through the TFNA aiming arm properly, which may cause surgical delay. · FDA Class II
Units
2,064 units
Sold at
Nationwide Distribution to AK,AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, and WY
Description
DePuy Synthes is initiating a voluntary medical device recall due to Blade/Screw Guide Sleeves (Part # 03.037.017) may have been manufactured with a slightly larger than specified outer diameter which could prevent the sleeve from going through the TFNA aiming arm properly, which may cause surgical delay.
Product
Depuy Synthes Trauma- Trochanteric Fixation Nail Advanced (TFNA) Blade/Screw Guide Sleeve, Surgical Instrument Part Number ID 03.037.017, Class I exempt Lot Numbers: 9066399; 9066435; 9409790; 9519308; 9620447; 9301183; 9409792; 9519303; 9641077; 9301184; 9409791; 9542998; 9641078; 9327510; 9422885; 9554330; 9651850; 9340629; 9351396; 9554329; 9604307; 9347642; 9428332; 9533341; 9661610; 9339162; 9439269; 9574931; 9734868; 9347641; 9485532; 9569618; 9704087; 9351397; 9485534; 9569617; 9734876; 9359540; 9494322; 9590744; 9661608; 9359339; 9501935; 9590747; 9371095; 9505477; 9604306 These devices are specific to the Trochanteric Fixation Nail Advanced (TFNA) system. The TFNA Proximal Femoral Nailing System is for the treatment of intramedullary fixation of proximal femoral fractures
Manufacturer(s)
Synthes (USA) Products LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →