Home / Recalls / Medical Devices
Synthes (USA) Products LLC recalls LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Ret
Recalled October 29, 2025 · FDA recall Z-0229-2026 · Synthes (USA) Products
Hazard
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm. · FDA Class II
Units
122 units
Sold at
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
Description
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Product
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Manufacturer(s)
Synthes (USA) Products LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →