RecallWatchUSA

Home / Recalls / Medical Devices

Synthes (USA) Products LLC recalls STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants dur

Recalled June 5, 2024 · FDA recall Z-1916-2024 · Synthes (USA) Products

Hazard
Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay · FDA Class II
Units
261 units
Sold at
Worldwide distribution - US Nationwide and the country of Canada.

Description

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

Product
STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010
Manufacturer(s)
Synthes (USA) Products LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov