Home / Recalls / Medical Devices
Synthes (USA) Products LLC recalls TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization
Recalled March 22, 2023 · FDA recall Z-1247-2023 · Synthes (USA) Products
Hazard
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length · FDA Class II
Units
5 units
Sold at
Worldwide - US Nationwide distribution in the states of MS, OR, TX and the countries of Canada, Japan.
Description
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Product
TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
Manufacturer(s)
Synthes (USA) Products LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →