Sysmex America, Inc. recalls Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for a

Recalled February 12, 2025 · FDA recall Z-1113-2025 · Sysmex America

Hazard
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System. · FDA Class II
Units
32 units
Sold at
US Nationwide.

Description

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Product
Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.
Manufacturer(s)
Sysmex America, Inc.