RecallWatchUSA

Home / Recalls / Medical Devices

Sysmex America, Inc. recalls Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sy

Recalled November 14, 2012 · FDA recall Z-0255-2013 · Sysmex America

Hazard
There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias. · FDA Class II
Units
763 units
Sold at
Nationwide distribution: USA including Puerto Rico.

Description

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Product
Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #063-7591-7 The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.
Manufacturer(s)
Sysmex America, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Sysmex America, Inc. recalls Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sy | RecallWatchUSA