Home / Recalls / Medical Devices
Sysmex America, Inc. recalls TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running s
Recalled January 28, 2026 · FDA recall Z-1136-2026 · Sysmex America
Hazard
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift. · FDA Class II
Units
44 units
Sold at
US Nationwide.
Description
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
Product
TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
Manufacturer(s)
Sysmex America, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →