Home / Recalls / Medical Devices
Talladium Inc recalls Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UD
Recalled September 16, 2020 · FDA recall Z-2943-2020 · Talladium
Hazard
The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range. · FDA Class II
Units
39 pieces
Sold at
U.S. Nationwide distribution including in the states of CA, FL, KS, LA, MO, and PA. No OUS Consignees.
Description
The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range.
Product
Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - Product Usage: are intended for CAD/CAM fabrication of all-ceramic (no metal) dental restorations. Luminesse ZR Blanks are partially sintered Yttria stabilized zirconia blanks for use as CAD/CAM milling blanks.
Manufacturer(s)
Talladium Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →