Tec Com Gmbh recalls TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product U

Recalled November 20, 2019 · FDA recall Z-0362-2020 · Tec Com

Hazard
The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power. · FDA Class II
Units
5 units
Sold at
US Nationwide distribution in the state of UT.

Description

The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.

Product
TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System TECOTHERM NEO is designed for controlled cold & heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective.
Manufacturer(s)
Tec Com Gmbh
Made in
Germany