Technidata S.A. recalls TD-Synergy. Laboratory Information System.

Recalled May 29, 2013 · FDA recall Z-1345-2013 · Technidata S.A

Hazard
Technidata has discovered that using the F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request. · FDA Class II
Units
10
Sold at
Worldwide Distribution-USA (nationwide) including Washington DC and the states of California, Nevada, Kentucky, Alabama, Rhode Island, Pennsylvania, and Louisiana and the country of Italy.

Description

Technidata has discovered that using the F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request.

Product
TD-Synergy. Laboratory Information System.
Manufacturer(s)
Technidata S.A.
Made in
France