Technidata S.A. recalls TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use

Recalled July 24, 2013 · FDA recall Z-1743-2013 · Technidata S.A

Hazard
A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and · FDA Class III
Units
1 site
Sold at
US Distribution CA only.

Description

A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and

Product
TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System (L.I.S)
Manufacturer(s)
Technidata S.A.
Made in
France

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov