Home / Recalls / Medical Devices
Teleflex Medical Europe Ltd recalls Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2", REF 142542 Product Usage: general surgical scissors
Recalled July 4, 2018 · FDA recall Z-2287-2018 · Teleflex Medical Europe
Hazard
The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized. · FDA Class II
Units
573 units
Sold at
Worldwide Distribution - US Nationwide and the countries of Canada, India
Description
The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized.
Product
Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2", REF 142542 Product Usage: general surgical scissors
Manufacturer(s)
Teleflex Medical Europe Ltd
Made in
Ireland