Teleflex Medical recalls Percuvance(R) Percutaneous Shaft 29cm, PCVSH3 The Percutaneous Surgical System with 5mm attachments is indicated for

Recalled January 30, 2019 · FDA recall Z-0753-2019 · Teleflex Medical

Hazard
There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required. · FDA Class II
Units
1751 units
Sold at
Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Iran, Ireland, Italy, Japan, Lebanon, Netherlands, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, and the United Arab Emirates.

Description

There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.

Product
Percuvance(R) Percutaneous Shaft 29cm, PCVSH3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.
Manufacturer(s)
Teleflex Medical