Home / Recalls / Medical Devices
Terumo Cardiovascular Systems Corporation recalls 18 Fr 6mm Sft Flow Ang Wire W/L The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta dur
Recalled December 12, 2012 · FDA recall Z-0460-2013 · Terumo Cardiovascular Systems
Hazard
The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl · FDA Class II
Units
5760 total
Sold at
Worldwide Distribution - USA including AL , CA , CO , DC , DE , FL , GA , IL , IN , KS , KY , LA , MA , MI , MO , MS , MT , NC , ND , NE , NH , NY , OH , PA , TN , TX , VA , and WI. Internationally to Australia, United Arab Emirates (UAE), Singapore, Malaysia, BELGIUM, Japan, and CANADA.
Description
The presence of plastic flash was identified at the tip of certain lots of Sarns Soft-Flow Aortic Cannula. The plastic flash has the potential to detach and contribute to an adverse patient effect. While this potential failure has not been reported from the field as an observed defect, it was determined that the product does not meet specifications and this condition could result in an unacceptabl
Product
18 Fr 6mm Sft Flow Ang Wire W/L The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
Manufacturer(s)
Terumo Cardiovascular Systems Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →