RecallWatchUSA

Home / Recalls / Medical Devices

Terumo Cardiovascular Systems Corporation recalls CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressu

Recalled September 16, 2015 · FDA recall Z-2742-2015 · Terumo Cardiovascular Systems

Hazard
Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69. · FDA Class II
Units
4638
Sold at
Worldwide Distribution-US (nationwide) including DC and PR and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Australia, Belgium, Canada, Chile, Colombia, Hong Kong, Indonesia, Japan, Malaysia, Mexico, Philippines, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates.

Description

Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69.

Product
CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature.
Manufacturer(s)
Terumo Cardiovascular Systems Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Terumo Cardiovascular Systems Corporation recalls CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressu | RecallWatchUSA