RecallWatchUSA

Home / Recalls / Medical Devices

Terumo Cardiovascular Systems Corporation recalls Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188

Recalled September 1, 2021 · FDA recall Z-2350-2021 · Terumo Cardiovascular Systems

Hazard
intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs · FDA Class II
Units
1 unit OUS
Sold at
International distribution in the country of Panama.

Description

intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs

Product
Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
Manufacturer(s)
Terumo Cardiovascular Systems Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Terumo Cardiovascular Systems Corporation recalls Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188 | RecallWatchUSA