Home / Recalls / Medical Devices
Terumo Cardiovascular Systems Corporation recalls Tested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channel system f
Recalled November 21, 2012 · FDA recall Z-0279-2013 · Terumo Cardiovascular Systems
Hazard
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up · FDA Class II
Units
15 units
Sold at
Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI
Description
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up
Product
Tested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.
Manufacturer(s)
Terumo Cardiovascular Systems Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →