Home / Recalls / Medical Devices
Terumo Cardiovascular Systems Corporation recalls The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directl
Recalled June 9, 2021 · FDA recall Z-1789-2021 · Terumo Cardiovascular Systems
Hazard
Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices. · FDA Class II
Units
75 devices
Sold at
Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam
Description
Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.
Product
The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810
Manufacturer(s)
Terumo Cardiovascular Systems Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →