Home / Recalls / Medical Devices
The Anspach Effort, Inc. recalls Irrigation Clip 30; Intended to position and hold irrigation tubing during surgical procedures.
Recalled March 30, 2016 · FDA recall Z-1246-2016 · The Anspach Effort
Hazard
Assembled with the incorrect (smaller) front clips (component 44-0495 instead of 44.0498). · FDA Class II
Units
50 clips
Sold at
Distributed in the states of PA and MN, and the country of Switzerland.
Description
Assembled with the incorrect (smaller) front clips (component 44-0495 instead of 44.0498).
Product
Irrigation Clip 30; Intended to position and hold irrigation tubing during surgical procedures.
Manufacturer(s)
The Anspach Effort, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →