RecallWatchUSA

Home / Recalls / Medical Devices

The Binding Site Group, Ltd. recalls Human lgG4 Subclass kit for use on the Beckman Coulter AU" series This kit is intended for qualifying human IgG subcla

Recalled February 1, 2017 · FDA recall Z-1068-2017 · The Binding Site Group

Hazard
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results. · FDA Class II
Units
N/A
Sold at
Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.

Description

Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

Product
Human lgG4 Subclass kit for use on the Beckman Coulter AU" series This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents Product Code LK009.AU
Manufacturer(s)
The Binding Site Group, Ltd.
Made in
United Kingdom

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

The Binding Site Group, Ltd. recalls Human lgG4 Subclass kit for use on the Beckman Coulter AU" series This kit is intended for qualifying human IgG subcla | RecallWatchUSA