The Binding Site Group, Ltd. recalls VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016

Recalled March 18, 2020 · FDA recall Z-1469-2020 · The Binding Site Group

Hazard
A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency. · FDA Class II
Units
462 kits
Sold at
Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.

Description

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Product
VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.
Manufacturer(s)
The Binding Site Group, Ltd.
Made in
United Kingdom