RecallWatchUSA

Home / Recalls / Medical Devices

The Metrix Company recalls Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000

Recalled August 3, 2016 · FDA recall Z-2261-2016 · The Metrix

Hazard
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications. · FDA Class II
Units
18,366 cases
Sold at
Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom

Description

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

Product
Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 mL.
Manufacturer(s)
The Metrix Company

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

The Metrix Company recalls Container fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL, 1500 mL, 2000 mL, 3000 mL, or 4000 | RecallWatchUSA