RecallWatchUSA

Home / Recalls / Medical Devices

Theralase Inc. recalls TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.

Recalled February 26, 2020 · FDA recall Z-1326-2020 · Theralase

Hazard
Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification) · FDA Class III
Units
76 units
Sold at
Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.

Description

Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)

Product
TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.
Manufacturer(s)
Theralase Inc.
Made in
Canada

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov