Home / Recalls / Medical Devices
Thermo Fisher Scientific (Asheville) LLC recalls Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 10L GP Bath Model Number: FSGPD10
Recalled July 3, 2019 · FDA recall Z-1866-2019 · Thermo Fisher Scientific (Asheville)
Hazard
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user. · FDA Class II
Units
5 units
Sold at
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Description
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Product
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 10L GP Bath Model Number: FSGPD10
Manufacturer(s)
Thermo Fisher Scientific (Asheville) LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →