RecallWatchUSA

Home / Recalls / Medical Devices

Thor Photomedicine Ltd recalls NovoTHOR Gen 2.5 XL whole body red light therapy bed, model S2184.

Recalled December 13, 2023 · FDA recall Z-0456-2024 · Thor Photomedicine

Hazard
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift. · FDA Class II
Units
26
Sold at
Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia. OUS (foreign) to countries of: Australia, Canada, Netherlands, New Zealand, Singapore, United Kingdom.

Description

Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.

Product
NovoTHOR Gen 2.5 XL whole body red light therapy bed, model S2184.
Manufacturer(s)
Thor Photomedicine Ltd
Made in
United Kingdom

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Thor Photomedicine Ltd recalls NovoTHOR Gen 2.5 XL whole body red light therapy bed, model S2184. | RecallWatchUSA