RecallWatchUSA

Home / Recalls / Medical Devices

Thoratec Switzerland GMBH recalls Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pu

Recalled October 24, 2018 · FDA recall Z-0103-2019 · Thoratec Switzerland

Hazard
Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection. · FDA Class II
Units
1569
Sold at
U.S.: OR, AZ, CA, ID, WA, WI, NY, MA, CT, NJ, ME, RI, IL, MI, TN, OH, IN, KY, IA, AL, FL, NC, GA, SC, VA, MD, DC, PA, WV, DE, MN, MT, NE, CO, NM, UT, AR, MO, MC, TX, OK, LA, MS, ND

Description

Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.

Product
Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.
Manufacturer(s)
Thoratec Switzerland GMBH
Made in
Switzerland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Thoratec Switzerland GMBH recalls Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pu | RecallWatchUSA