RecallWatchUSA

Home / Recalls / Medical Devices

Topcon Medical Systems, Inc. recalls Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained he

Recalled February 10, 2021 · FDA recall Z-0952-2021 · Topcon Medical Systems

Hazard
Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record · FDA Class II
Units
13 units
Sold at
AZ, MO, NC, NY, PA, TX

Description

Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record

Product
Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained healthcare professionals and clinicians to access archived multimedia data, forms and patient information.
Manufacturer(s)
Topcon Medical Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Topcon Medical Systems, Inc. recalls Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained he | RecallWatchUSA