Topcon Medical Systems, Inc. recalls The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and man
Recalled November 20, 2024 · FDA recall Z-0362-2025 · Topcon Medical Systems
Hazard
The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided · FDA Class II
Units
3144 units
Sold at
US Nationwide distribution.
Description
The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided
Product
The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices. It is intended for processing and displaying ophthalmic images and optical coherence tomography data. The IMAGEnet 6 Ophthalmic Data System uses the same algorithms and reference databases from the original data capture device as a quantitative tool for the comparison of posterior ocular measurements to a database of known normal subjects.
Manufacturer(s)
Topcon Medical Systems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →