Toshiba American Medical Systems Inc recalls Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F

Recalled June 7, 2017 · FDA recall Z-2109-2017 · Toshiba American Medical Systems

Hazard
It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on. · FDA Class II
Units
US - 70 systems
Sold at
US Distribution to the states of :LA, IL, NY, FL, MA, TN, MO, GA, CA, DE, MO, AL, ID, MN, TX, CT, PA, MS, OH, NC, NV, W VA, WA, AR, NJ, MD, OK, IN and NH.

Description

It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.

Product
Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F
Manufacturer(s)
Toshiba American Medical Systems Inc