Toshiba American Medical Systems Inc recalls Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.

Recalled November 20, 2013 · FDA recall Z-0201-2014 · Toshiba American Medical Systems

Hazard
Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor. · FDA Class II
Units
18 units
Sold at
USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.

Description

Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.

Product
Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
Manufacturer(s)
Toshiba American Medical Systems Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov