Toshiba American Medical Systems Inc recalls Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoros

Recalled June 11, 2014 · FDA recall Z-1714-2014 · Toshiba American Medical Systems

Hazard
Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop. · FDA Class II
Units
2
Sold at
Nationwide Distribution including CA and IN

Description

Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.

Product
Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.
Manufacturer(s)
Toshiba American Medical Systems Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov