Toshiba American Medical Systems Inc recalls Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

Recalled April 17, 2013 · FDA recall Z-1068-2013 · Toshiba American Medical Systems

Hazard
Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal. · FDA Class II
Units
1 unit
Sold at
Distributed in the state of MD.

Description

Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.

Product
Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
Manufacturer(s)
Toshiba American Medical Systems Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov