Home / Recalls / Medical Devices
Tosoh Smd Inc recalls AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyz
Recalled August 7, 2013 · FDA recall Z-1830-2013 · Tosoh Smd
Hazard
Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers. · FDA Class II
Units
11,309 bags
Sold at
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. Internationally to Canada, Japan, Mexico, Chile, and Uruguay.
Description
Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.
Product
AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.
Manufacturer(s)
Tosoh Smd Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →