Trek Diagnostic Systems recalls VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually p
Recalled September 12, 2012 · FDA recall Z-2323-2012 · Trek Diagnostic Systems
Hazard
Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database. · FDA Class III
Units
333 units
Sold at
Worldwide Distribution - USA including AL, AR, AZ, CA, CO , CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SD, TN, TX, VA, VT, WA, WI, WV, and WY. Internationally to the countries of Bahamas, Bangladesh, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, India, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Pakistan, Peru, Philippines, Portugal, Puerto Rico, Russian Federation, South Africa, Spain, Thailand, Turkey, and United Kingdom
Description
Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database.
Product
VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually packaged using a standard CD plastic case and placed into a standard CD mailer The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile body fluids.
Manufacturer(s)
Trek Diagnostic Systems
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →